
1-8
IMSBR10-E1
• Delay High Limit Alarm (T)
An alarm occurs when the measured value remains above the alarm value for a specified time period (alarm
delay period).
• Delay Low Limit Alarm (t)
An alarm occurs when the measured value remains below the alarm value for a specified time period (alarm
delay period).
Delay High Limit Alarm Example (T is the specified delay)
Alarm release
Alarm occurrence
T
X1
X2
X3
X4
T1
Measured value
Alarm value
• Alarm does not occur at T1, because the time period is shorter than the specified alarm delay period (T).
• The measured value exceeds the alarm value at time X2, and the alarm occurs at time X3 at which the
specified alarm delay period elapses (the time when the alarm occurs is the time at X3).
• The measured value falls below the alarm value at time X4, and the alarm is released.
Note
• The alarm detection operation is reset when a power failure occurs. The operation restarts after the
power recovers.
• If the alarm value is changed while a delay alarm is occurring, the alarm is released if the new alarm
value does not meet the alarm condition.
Alarm Hysteresis
Hysteresis can be specified to the values for activating and releasing the alarm. The hysteresis applies only to
high limit alarm (H) and low limit alarm (L). The hysteresis width can be set in the range of 0.0% (Off) to 1.0% of
the recording span in 0.1 steps. The setting applies to all high limit alarms and low limit alarms. By default, the
hysteresis width is set to 0.5%.
Alarm
value
Alarm release
Measured value
Alarm occurrence
High limit alarm
Low limit alarm
Measured
value
Alarm release
Hysteresis
(1% or less)
Alarm value
Alarm occurrence
<Related Topics>
Setting alarms: Section 5.2
Setting the alarm delay function: Section 7.15 and 6.10
Setting the alarm hysteresis: Section 7.1
1.3 Alarms
Summary of Contents for SBR-EW180
Page 1: ...IMSBR10 E1 SBR EW180 Recorder User s Manual...
Page 2: ......
Page 14: ......
Page 233: ...3 2002...