System installation
4-3
Figure 4.1 – RCU10 status display
Connect the PC to the RCU10 that is to be configured, using the PC cable provided.
Click the
Receive
button on the RCU CS screen. The connected RCU10 unit
should be detected with a default address of 1.
Select the
Compensator Address
function from the
Configure
menu. A
confirmation dialogue will appear, displaying the current RCU10 address setting.
Select
Yes
to change the address.
Enter the new system address in the window displayed. The addresses can be in
the range 1 to 6.
NOTE:
Each unit
must
have a unique address before connection as part of a network, for
example a six-axis system must have RCU10s addressed as 1, 2, 3, 4, 5 and 6. It is
important to ensure that compensators are numbered sequentially. They may be
connected in any order, however the physical end units must be identified in the software
for correct termination.
Once you have acknowledged the new address, the RCU CS will reset the RCU10
and re-establish communications. The connected RCU10 will now register with its
new address.
Repeat this process of setting network addresses for
all
RCU10 units in the system.
Button bar
Menu bar
Status bar
Serial
port
status
RCU10
connection
status
Current
RCU CS
operating
mode
Current
access
level
Summary of Contents for RCU10
Page 1: ...Installation and user s guide M 9904 1122 09 A RCU10 quadrature compensation unit...
Page 11: ...viii Contents This page is intentionally left blank...
Page 23: ...1 12 System overview This page is intentionally left blank...
Page 39: ...2 16 System design...
Page 49: ...3 6 Kit configuration and part identification This page is intentionally left blank...
Page 67: ...4 18 System installation This page is intentionally left blank...
Page 91: ...A 6 RCU10 system specifications This page is intentionally left blank...
Page 147: ...E 8 Commissioning tests This page is intentionally left blank...
Page 175: ...F 28 Extended capability This page is intentionally left blank...
Page 189: ...G 14 Reference This page is intentionally left blank...
Page 193: ...H 4 Test records This page is intentionally left blank Cut here Cut here...
Page 195: ...H 6 Test records This page is intentionally left blank Cut here Cut here...