Remote Control - Commands
R&S
®
NRP2
317
User Manual 1173.9157.02 ─ 03
SYSTem:COMMunicate:INET:MODE
<mode>
Selects the mode for transmitting the special control sequences via ethernet according
to draft IEEE 1174.
Parameter
Mode
0
no control sequences (abbr.: 1174.0 mode)
1
the IEC/IEEE 488.1 mode (abbr.: 1174.1 mode)
2
the IEC/IEEE 488.2 mode (abbr.: 1174.2 mode)
Parameters:
<state>
0 | 1 | 2
*RST:
This setting is not changed by *RST.
SYSTem:COMMunicate:INET[:SELF]:ADDRess
<ip1>,<ip2>,<ip3>,<ip4>
Sets the IP address of the R&S
NRP2.
Note: This command cannot be used, while VXI11 or Ethernet is in use.
Note: Choosing an invalid IP-Address may disturb the traffic on your LAN. If you are not
sure how to configure these settings, please ask your network administrator.
Parameters:
<ip1>,<ip2>,<ip3>,<ip4>
int_values
The four parameter form the IP address <ip1>.<ip2>.<ip3>.<ip4>.
For example, IP address 104.0.0.255 is entered as
SYSTem:COMMunicate:INET:ADDRess 108,0,0,255
.
Range:
0...255 for each part of the IP address
*RST:
This setting is not changed by *RST.
Usage:
Event
SYSTem:COMMunicate:INET[:SELF]:GATEway:ADDRess
<gwip1>,<gwip2>,<gwip3>,<gwip4>
Sets the gateway address of your local subnet.
Note: This command cannot be used, while VXI11 or Ethernet is in use.
Note: Choosing an invalid Gateway-Address may disturb the traffic on your LAN. If you
are not sure how to configure these settings, please ask your network administrator.
Parameters:
<gwip1...4>
int_value
Range:
0...255
*RST:
This setting is not changed by *RST.
Usage:
Event
SYSTem
Summary of Contents for NRP2
Page 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Page 20: ......
Page 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Page 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Page 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Page 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Page 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Page 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Page 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...