Table 5-2: Synchronization with *OPC, *OPC? and *WAI
Command
Effect
Programming of controller
*OPC
Sets the Operation Complete bit in the ESR
after all commands sent prior to *OPC have
been processed.
Set bit 0 in the ESE of the R&S
NRP2.
Set bit 5 in the SRE of the R&S
NRP2.
Wait for service request (SRQ).
*OPC?
Writes 1 in the output buffer after all com-
mands sent prior to
*OPC?
have been pro-
cessed.
IEC/IEEE bus: the R&S
NRP2 is addressed
as a talker.
*WAI
Commands can be sent but they are only exe-
cuted after completion of all commands sent
prior to
*WAI
.
Dividing commands timewise before/after
*WAI
serializes the command sequence.
*WAI
is sent directly after the command that
should be completed before other com-
mands can be executed.
5.9 Automatic Settings for Manual Selection of a Measure-
ment Mode
When the measurement mode of a sensor is manually selected ("Sensor"
→
"Mode"),
automatic settings are made for the trigger system and timing, which cannot be changed
by manual control. This considerably simplifies operation. These settings are not changed
when the measurement mode is selected by remote control (
SENSe:FUNCtion
) and
have to be set to the desired value by appropriate SCPI commands.
The following table lists the respective setting capabilities and the values that are auto-
matically set in manual control.
Table 5-3: Automatic settings for manual selection of a measurement mode
SCPI command
Value
Remark
Mode Cont Avg
TRIGger[1..4]:SOURce
IMMediate
Cannot be modified in manual mode.
TRIGger[1..4]:SLOPe
-
Unchanged, is irrelevant.
TRIGger[1..4]:COUNt
1
Cannot be modified in manual mode.
TRIGger[1..4]:DELay
-
Unchanged, is irrelevant.
TRIGger[1..4]:HOLDoff
1 NS
Cannot be modified in manual mode.
TRIGger[1..4]:LEVel
-
Unchanged, is irrelevant.
[SENSe[1..4]]:TIMing:EXCLude:
STARt
-
Unchanged, is irrelevant.
[SENSe[1..4]]:TIMing:EXCLude:STOP
-
Unchanged, is irrelevant.
INITiate[1..4]:CONTinuous
ON
Cannot be modified in manual mode.
Mode Burst Av
TRIGger[1..4]:SOURce
-
Unchanged, is irrelevant.
Automatic Settings for Manual Selection of a Measurement Mode
Summary of Contents for NRP2
Page 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Page 20: ......
Page 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Page 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Page 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Page 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Page 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Page 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Page 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...