Instrument Functions
R&S
®
NRP2
67
User Manual 1173.9157.02 ─ 03
The more modulation periods fit into an acquisition interval, the less critical it is whether
N is an integer or not. With smoothing on, about five periods are enough to reduce fluc-
tuations caused by modulation to an acceptable level - if N is not an integer. With more
than nine periods, the fluctuations are imperceptible. With smoothing off, the situation is
much less favorable as instead of just five periods as many as 300 are required; the
fluctuations only completely disappear with more than 3000 periods.
It is not useful to select the aperture time longer than required, since it increases the noise
component of the power sensor (
chapter 4.2.1.7, "Background information on power
SCPI command:
Sampling rate
Sets the sampling rate. With the multipath power sensors R&S
NRP-
Z11/21/22/23/24/28/31/32, the sampling rate can be set to two different values to prevent
aliasing effects for particular types of modulation signal. Aliasing can occur because the
sampling frequency is located within the video bandwidth, which means that spectral
components near the sampling frequency can cause beating effects. With changing the
sampling rate, the beating effects usually disappear.
SCPI command:
Duty Cycle
Using the duty cycle correction, the average power of RF bursts is calculated from the
average power of the whole signal. Essentially, the average power of the whole signal is
divided by the set duty cycle.
If thermal power sensors or R&S
NRP-Z91/98 multipath power sensors are used, this
procedure is the only way to determine the power of pulsed signals.
If multipath power sensors R&S
NRP-Z11/21/22/23/24/28/31/32 are used, this method
must be employed if the bursts are so short that their power cannot be measured in the
"Burst Av" or "Timegate Av" modes.
SCPI command:
Smoothing of modulated signals
To reduce result fluctuations caused by modulation, this parameter should always be
activated if the aperture time cannot be exactly adjusted to the modulation period. If the
aperture time is only 5 to 9 times larger than a modulation period, the display fluctuations
are sufficiently reduced. With smoothing off, 300 to 3000 periods would be required to
obtain the same effect (
SCPI command:
Data acquisition and parameters
Summary of Contents for NRP2
Page 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Page 20: ......
Page 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Page 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Page 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Page 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Page 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Page 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Page 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...