Instrument Functions
R&S
®
NRP2
85
User Manual 1173.9157.02 ─ 03
Source
Γ
Magnitude
Magnitude of the source reflection coefficient
Γ
.
SCPI command:
Source
Γ
Phase
Phase angle of the source reflection coefficient
Γ
in degrees.
SCPI command:
4.2.6 Trigger settings
The trigger system is required for the "Trace & Statistics", "Timeslot Av", "Timegate Av",
and "Burst Av" measurement modes.
► Configure the trigger system in the "Trigger" dialog box which is opened via the
"Sensor" menu, or via the control panel of the respective trace mode.
Trigger dialog box, page1
The first page covers the parameters for defining a trigger mode, trigger source, and
trigger level.
Fig. 4-20: Trigger dialog box, page1
A | B | C | D
Tabs for selecting the sensor.
Note:
In remote-control commands, select the sensor with the suffix in command
SENSe<[1]..4>:...
, i.e. choose suffix
1
for channel A, channel
2
for B, and so on.
Mode
Selects the trigger mode.
Selections "Single" and "Normal" are only available in "Trace&Statistics" measurement
mode.
Data acquisition and parameters
Summary of Contents for NRP2
Page 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Page 20: ......
Page 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Page 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Page 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Page 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Page 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Page 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Page 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...