Remote Control - Commands
R&S
®
NRP2
223
User Manual 1173.9157.02 ─ 03
...............................................233
...........................................233
.....................................233
...........................................................234
CALCulate<[1]..8>:COUNt:DATA?
Queries the number of measured values that are included when the mean value and
standard deviation are calculated (in the above formulas referred to as N, see
and
).
Usage:
Query only
CALCulate<[1]..8>:DATA?
With this command the current output value of a calculate block can be queried.
The unit is the current unit of the measured result.
Possible Units:
W | DBM | DBUV | DPCT | DB | O
Usage:
Query only
CALCulate<[1]..8>:EXTRemes:RESet
The R&S
NRP2 stores the maximum and minimum values for each calculate block. The
current measured value is stored with
CALC[1..8]:EXTR:RES
as the new minimum
and maximum value. The extreme values can be queried with
CALC[1..8]:MIN:DATA?
and
CALC[1..8]:MAX:DATA?
.
Usage:
Event
CALCulate<[1]..8>:FEED<[1]|2>
<feed>
The sensors connected with the calculate block via the primary and secondary channel
can yield differently processed data: peak value, average value, peak-to-average value
or even a measurement sequence, as is the case in the Trace mode. The
CALC:FEED
determines which of this data is to be processed by the calculate block. Depending on
the measurement mode, the following settings are possible:
CALCulate (Configuration of Calculate Blocks)
Summary of Contents for NRP2
Page 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Page 20: ......
Page 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Page 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Page 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Page 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Page 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Page 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Page 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...