Instrument Functions
R&S
®
NRP2
80
User Manual 1173.9157.02 ─ 03
acquisitions have been evaluated. The oldest result is taken out of the averaging filter
and a new average value is calculated.
The averaging number can be selected automatically or it can be manually set to a fixed
value. As a starting point, automatic filtering should always be used. If the automatically
selected filter setting proves to be not adequate, the averaging number can be
in(de)creased manually.
Two autofilter operating modes are provided:
●
"Normal"
The "Normal" mode finds a balance between measurement time and display noise.
As an additional parameter, display resolution can be used to adjust the filter setting.
With increasing resolution, i.e. an increasing number of digits in the measurement
results, the averaging number also increases.
●
"Fixed Noise"
In the "Fixed Noise" mode, the averaging number is selected so that the sensors
intrinsic noise (2 standard deviations) does not exceed the specified "Noise Con-
tent". To avoid very long settling times when the power is low, the averaging number
can be limited with the "Max Settling Time" parameter. If the display noise then
exceeds the set noise content, the symbol "S/N" is displayed.
4.2.3.1
Filter dialog box
Fig. 4-16: Filter dialog box
A | B | C | D
Tabs for selecting the sensor.
Note:
In remote-control commands, select the sensor with the suffix in command
SENSe<[1]..4>:...
, i.e. choose suffix
1
for channel A, channel
2
for B, and so on.
Auto
Toggles between automatic and manual filter setting.
SCPI command:
Avg No
Sets the averaging number manually.
Data acquisition and parameters
Summary of Contents for NRP2
Page 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Page 20: ......
Page 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Page 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Page 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Page 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Page 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Page 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Page 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...