Remote Control - Commands
R&S
®
NRP2
204
User Manual 1173.9157.02 ─ 03
command, see
chapter 6.4, "High-Level Measurement Commands (CONFigure, MEA-
The individual commands are described in alphabetical order in the following chapters.
The high-level measurement commands are put first.
TRIGger
SENSe
OUTPut
DISPlay
SYSTem
MEMory
(CONFigure, FETCh, READ, MEASurement)
High-level measurement commands
Sensor referenced
command systems
Measured
values
Manual control
Remote control
- IEC-Bus
- Ethernet
Information flow
STATus
UNIT
CALCulate
Fig. 6-1: Overview of SCPI command systems and their mutual interaction
The figure above gives an overview of the SCPI command systems of the R&S
NRP2.
The measured values are forwarded to the SENSe block where they are processed and
handed to the
CALCulate
blocks. The values from the
SENSe
block are not immediately
made available to the user, but only after they have been processed in the
CALCulate
blocks. In the
CALCulate
blocks, the values of several sensors can be processed
together before they are output. High-level commands may be used to configure
SENSe
and
CALCulate
blocks with a single command. This may considerably simplify
operation of the R&S
NRP2. The frequency-dependent correction tables for the mea-
sured values are managed by the
MEMory
command system.
Table 6-4: Overview of command systems and their function
System
Function
ABORt
Stops the current measurement
CALCulate
Processes the measured data
CALibrate
Calibrates the R&S
NRP2
CONFigure
Specifies measurement settings (high-level commands)
DISPlay
Configures the reading on the R&S
NRP2 display.
FETCh
Retrieves measurement results from the running measurement (high-level com-
mands)
FORMat
Sets the data format on the remote control interfaces.
INITiate
Starts measurement
MEASurement
Configures and starts a measurement (high-level commands)
MEMory
Manages frequency correction tables
OUTPut
Switches the test generator on/off and calibrates it.
Overview of SCPI Command Systems
Summary of Contents for NRP2
Page 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Page 20: ......
Page 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Page 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Page 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Page 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Page 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Page 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...