[SENSe<[1]..4>]:CORRection:FDOFfset[:INPut][:MAGNitude]?
Queries the frequency dependent offset. The frequency-dependent measured-value cor-
rection is set by selecting a table with the
[SENSe<[1]..4>]:CORRection:FDOTable[:SELect]
and
[SENSe<[1]..4>]:CORRection:FDOTable:STATe ON
commands. The respective
frequency-dependent correction factor is determined by means of the active tables (if
necessary by interpolation) and can be queried with
[SENSe<[1]..4>]:CORRection:FDOFfset[:INPut][:MAGNitude]?
(fre-
quency-dependent offset).
Unit:
DB
Usage:
Query only
[SENSe<[1]..4>]:CORRection:FDOTable[:SELect]
<table_name>
The name of the current frequency-dependent offset table can be set or queried by the
[SENSe<[1]..4>]:CORRection:FDOTable[:SELect][?]
command. The
R&S
NRP2 can manage up to ten tables (
MEMory
commands). The frequency-dependent
offset tables consist of two columns; the first containing the frequencies, the second the
associated correction values. When the R&S
NRP2 is informed about the signal fre-
quency with command
[SENSe<[1]..4>]:FREQuency[:CW|:FIXed]
<float_value>
, an appropriate correction factor is determined by means of the table.
If the table was activated with
[SENSe<[1]..4>]:CORRection:FDOTable:STATe[?] ON
, the measurement
result obtained by the sensor is multiplied by this factor.
Error messages:
-256
"File name not found"
: An invalid <string> parameter was transmitted.
Parameters:
<table_name>
string
*RST:
Settings are not changed in case of *RST.
[SENSe<[1]..4>]:CORRection:FDOTable:STATe
<state>
Activates and deactivates the current frequency-dependent offset table. The current table
is selected with the
[SENSe<[1]..4>]:CORRection:FDOTable[:SELect]
com-
mand. Only one table can be active at a time.
Parameters:
<state>
ON | OFF
*RST:
OFF
[SENSe<[1]..4>]:CORRection:SPDevice:SELect
<NR1>
Several S-parameter tables can be stored in a sensor.
SENSe (Sensor Commands)
Summary of Contents for NRP2
Page 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Page 20: ......
Page 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Page 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Page 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Page 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Page 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Page 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Page 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...