Remote Control - Commands
R&S
®
NRP2
292
User Manual 1173.9157.02 ─ 03
SERVice<[1]..4>:RCOunt
<nr1>
SERVice<[1]..4>:RESult
<nr1>
Allows simulated measured values to be specified for test purposes.
SERV:RCO
specifies
the number of
SERV:RES
pairs to follow. The first
SERV:RES
command specifies the
measured value in watts, the second
SERV:RES
command determines how often the
measured value should be obtained. The measurement sequences defined in this way
are repeated until the sensor is disconnected from the R&S
NRP2 and reconnected.
Parameters:
<nr1>
int_value
*RST:
none. The simulation mode is not exited in the case
of *RST. To again obtain real measured values,
SERV:RCO 0 must be set.
Example:
The command sequence
SERV2:RCO 2
SERV2:RES 3
SERV2:RES 2
SERV2:RES 4
SERV2:RES 1
makes sensor 2 return the measured values 3 W, 3 W, 4 W (, 3
W, 3 W, 4W, ....) from now on.
SERVice<[1]..4>:SERRors
<state>
The sensors signal an error to the R&S
NRP2 if samples are lost while a signal is sampled
(sampling), as may be the case with high sampling rates. In such a case, the measure-
ment is less accurate. Since this error is signaled as an SCPI error, smooth operation of
a remote-control program may be disrupted. For this reason, sampling errors can be
suppressed by means of
SERVice<[1]..4>:SERRors OFF
.
Parameters:
<state>
ON | OFF
*RST:
The R&S NRP2 BIOS setting is adopted.
SERVice<[1]..4>:UNLock
<password>
Before other SERVice system commands are accepted, they must be enabled with
SERV:UNL 1234
.
Parameters:
<password>
1234
6.13 STATus
The status registers of the R&S
NRP2 can be read and configured with commands of the
STATus
system.
STATus
Summary of Contents for NRP2
Page 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Page 20: ......
Page 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Page 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Page 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Page 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Page 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Page 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Page 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...