Remote Control - Commands
R&S
®
NRP2
281
User Manual 1173.9157.02 ─ 03
Parameters:
<rlevel>
float_value
Range:
depending on sensor
*RST:
depending on sensor
Default unit: DBM
[SENSe<[1]..4>]:STATistics:TIME
< time_interval>
Determines, together with the
SENS:STAT:OFFS
command, the time interval in which
the power for the statistical evaluation is measured.
SENS:STAT:TIME
determines the
length of the interval (see also
Parameters:
< time_interval>
float_value
Range:
depending on sensor
*RST:
depending on sensor
Default unit: S
[SENSe<[1]..4>]:TIMing:EXCLude:STARt
<exclude_start>
Sets a time that is to be excluded at the beginning of the integration period.
Integration period
EXCLude:STARt
EXCLude:STOP
Power
Time
Fig. 6-15: Effect of commands SENS<[1]..4>:TIM:EXCL:STAR and :STOP
Parameters:
<exclude_start>
float_value
Range:
depending on sensor
*RST:
depending on sensor
Default unit: S
[SENSe<[1]..4>]:TIMing:EXCLude:STOP
<exclude_stop>
Sets a time that is to be excluded at the end of the integration period. See
Effect of commands SENS<[1]..4>:TIM:EXCL:STAR and :STOP
.
SENSe (Sensor Commands)
Summary of Contents for NRP2
Page 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Page 20: ......
Page 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Page 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Page 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Page 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Page 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Page 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Page 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...