Remote Control - Commands
R&S
®
NRP2
331
User Manual 1173.9157.02 ─ 03
Parameters:
<state>
BUS | EXTernal | HOLD | IMMediate | INTernal
BUS
The trigger event is initiated by
TRIG:IMM
or
*TRG
. In this case,
the other trigger setting are meaningless.
EXTernal
Triggering is performed with an external signal applied to the trig-
ger connector. The
TRIG:SLOP
command determines whether
the rising or the falling edge of the signal is to be used for trigger-
ing. Waiting for a trigger event can be skipped by
TRIG:IMM
.
HOLD
A measurement can only be triggered when the command
TRIG:IMM
is executed.
IMMediate
The sensor does not remain in the WAIT_FOR_TRG state but
immediately changes to the MEASURING state.
INTernal
The sensor determines the trigger time by means of the signal to
be measured. When this signal exceeds (
TRIG:SLOP POS
) or fall
short of (
TRIG:SLOP NEG
) the power set by
TRIG:LEV
, the mea-
surement is started after the time set by
TRIG:DEL
. Similar to
TRIG:SOUR EXT
, waiting for a trigger event can also be skipped
by
TRIG:IMM
.
*RST:
depending on sensor
TRIGger<[1]..4>:SYNChronize[:STATe]
<state>
If set to
ON
, the command blocks the external trigger bus as long as the sensor is not in
WAIT_FOR_TRIGGER state. It is usually used in combination with command
and makes certain, that a new measurement will only be started after all
sensors have completed their last measurement.
Parameters:
<state>
ON | OFF
*RST:
depending on sensor
6.16 UNIT
Commands of the UNIT system
............................................................................................332
UNIT<[1]..8>:POWer
<output_unit>
Selects the output unit for the measured power values.
UNIT
Summary of Contents for NRP2
Page 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Page 20: ......
Page 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Page 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Page 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Page 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Page 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Page 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Page 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...