Remote Control - Commands
R&S
®
NRP2
230
User Manual 1173.9157.02 ─ 03
CALCulate<[1]..8>:MAXimum:DATA?
CALCulate<[1]..8>:MINimum:DATA?
Queries the maximum/minimum of all measured values that previously occurred in this
calculate block.
The limit value is set to the current measured value when
●
the device is switched on
●
a reset is performed (
*RST
)
●
the value is explicitly set to the current measured value using the command
CALCulate[1..8]:EXTRemes:RESet
.
The unit is the current unit of the measured result.
Possible Units:
W | DBM | DBUV | DPCT | DB | O
Usage:
Query only
CALCulate<[1]..8>:MEAN:DATA?
Queries the mean value of all measured values of the calculate block. The mean value
is reset when the auxiliary values are reset. The following formula applies:
N
i
i
x
N
x
1
1
Fig. 6-5: Mean value of all measured values
The unit is the current unit of the measured result.
Possible Units:
W | DBM | DBUV | DPCT | DB | O
Usage:
Query only
CALCulate<[1]..8>:PTPeak:DATA?
Queries the peak-to-peak distance (maximum to minimum) of the measured values of
the calculate block.
The unit is the current unit of the measured result.
Possible Units:
W | DBM | DBUV | DPCT | DB | O
Usage:
Query only
CALCulate<[1]..8>:RELative[:MAGNitude]
<value>
Determines a value that is used as a divisor (logarithmic subtraction) for all measured
values of the calculate block in the
CALC:REL:STAT ON
state. The unit of this value is
equal to the unit of the output value of the calculate block and is defined according to the
following table:
CALCulate (Configuration of Calculate Blocks)
Summary of Contents for NRP2
Page 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Page 20: ......
Page 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Page 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Page 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Page 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Page 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Page 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Page 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...