Instrument Functions
R&S
®
NRP2
66
User Manual 1173.9157.02 ─ 03
"Burst Av"
The
Burst Av
erage mode is used to measure the average burst power
of pulsed signals. The acquisition interval is automatically derived from
the signal (see
chapter 4.2.1.3, "Measuring the average power of burst
SCPI command:
4.2.1.2
Continuous average-power measurements
The "Cont Av" mode is the preferred measurement method if the measurement is not to
be, or cannot be, synchronized with a specific signal event.
This is the only available measurement mode for thermal power sensors because they
are too slow for the other measurement modes.
► Select "Cont Av" in the "Mode..." dialog box opened via the "Sensor" menu.
Fig. 4-5: Mode dialog box / Cont Av
Aperture
Defines the aperture time (size of the acquisition interval). In the manual operating mode,
the sensor-specific default setting combined with active smoothing (see below) is usually
adequate. Another value, generally larger, is helpful if the result exhibits fluctuations due
to modulation. In particular with very low frequency modulation, we recommend that you
adjust the aperture time exactly to the modulation period, in order to obtain stable read-
ings:
Optimal aperture time
Smoothing
Off
N x ( modulation period / 2)
On
N x ( modulation period * 2)
N = 1, 2, 3, …
I.e. the minimum theoretical measurement time can only be obtained with smoothing
turned off.
Data acquisition and parameters
Summary of Contents for NRP2
Page 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Page 20: ......
Page 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Page 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Page 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Page 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Page 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Page 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Page 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...