Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
257
User Manual 1173.9240.02 ─ 23
TRIGger[:SEQuence]:LEVel:AM[:ABSolute]
...................................................................... 261
...................................................................................... 261
...................................................................................... 261
.......................................................................................... 262
........................................................................................262
TRIGger[:SEQuence]:TIME:RINTerval
............................................................................. 264
TRIGger[:SEQuence]:BBPower:HOLDoff
<Period>
This command defines the holding time before the baseband power trigger event.
The command requires the optional Digital Baseband Interface or the optional Analog
Baseband Interface.
Note that this command is maintained for compatibility reasons only. Use the
TRIGger[:SEQuence]:IFPower:HOLDoff
on page 258 command for new remote
control programs.
Parameters:
<Period>
Range:
150 ns to 1000 s
*RST:
150 ns
Example:
TRIG:SOUR BBP
Sets the baseband power trigger source.
TRIG:BBP:HOLD 200 ns
Sets the holding time to 200 ns.
TRIGger[:SEQuence]:DTIMe
<DropoutTime>
Defines the time the input signal must stay below the trigger level before a trigger is
detected again.
For input from the Analog Baseband Interface (R&S
FSW-B71) using the baseband
power trigger (BBP), the default drop out time is set to 100
ns to avoid unintentional
trigger events (as no hysteresis can be configured in this case).
Parameters:
<DropoutTime>
Dropout time of the trigger.
Range:
0 s to 10.0 s
*RST:
0 s
Manual operation:
See
TRIGger[:SEQuence]:HOLDoff[:TIME]
<Offset>
Defines the time offset between the trigger event and the start of the sweep.
Parameters:
<Offset>
*RST:
0 s
Example:
TRIG:HOLD 500us
Manual operation:
See
Configuring the Measurement