Configuration
R&S
®
FSW-K7
82
User Manual 1173.9240.02 ─ 23
For the second alignment step, the connector must be disconnected from the REF
OUTPUT 640
MHZ connector and instead connected to the FSW B2000 ALIGNMENT
SIGNAL SOURCE connector on the R&S
FSW.
To continue the alignment, select the "Continue Alignment" button.
After the second alignment step has been completed successfully, a new dialog box is
displayed.
In order to switch from alignment mode to measurement mode, move the cable from
the FSW B2000 ALIGNMENT SIGNAL SOURCE back to the IF OUT 2
GHZ connec-
tor, so that it is then connected to the CH1 input on the oscilloscope.
If UNCAL is displayed, alignment was not yet performed (successfully).
Input and Frontend Settings