Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
207
User Manual 1173.9240.02 ─ 23
Usage:
Query only
SYSTem:COMMunicate:RDEVice:OSCilloscope:ALIGnment:DATE?
Returns the date of alignment of the IF OUT 2
GHZ to the oscilloscope for the optional
2
GHz bandwidth extension (R&S
FSW-B2000).
Return values:
<Date>
Returns the date of alignment.
Example:
SYST:COMM:RDEV:OSC:DATE?
//Result: 2014-02-28
Usage:
Query only
SYSTem:COMMunicate:RDEVice:OSCilloscope:IDN?
Returns the identification string of the oscilloscope connected to the R&S
FSW.
Return values:
<IDString>
Example:
SYST:COMM:RDEV:OSC:IDN?
//Result: Rohde&Schwarz,RTO,
1316.1000k14/200153,2.45.1.1
Usage:
Query only
Manual operation:
See
"TCPIP Address or Computer name"
SYSTem:COMMunicate:RDEVice:OSCilloscope:LEDState?
Returns the state of the LAN connection to the oscilloscope for the optional 2
GHz
bandwidth extension (R&S
FSW-B2000).
Return values:
<Color>
GREEN
Connection to the instrument has been established successfully.
GREY
Configuration state unknown, for example if you have not yet
started transmission.
RED
Connection to the instrument could not be established.
Check the connection between the R&S
FSW and the oscillo-
scope, and make sure the IP address of the oscilloscope has
been defined (see
Example:
SYST:COMM:RDEV:OSC:LEDS?
//Result: 'GREEN'
Usage:
Query only
Configuring the Measurement