Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
208
User Manual 1173.9240.02 ─ 23
SYSTem:COMMunicate:RDEVice:OSCilloscope:SRATe
<Rate>
Determines whether the 10
GHz mode (default) or 20
GHz mode of the connected
oscilloscope is used. The 20
GHZ mode achieves a higher decimation gain, but
reduces the record length by half.
Return values:
<Rate>
10 GHz | 20 GHz
No other sample rate values are allowed.
*RST:
10 GHz
Example:
TRAC:IQ:SRAT?
//Result: 100000000
TRAC:IQ:RLEN?
//Result: 3128
SYST:COMM:RDEV:OSC:SRAT 20GHZ
TRAC:IQ:SRAT?
//Result: 200000000
TRAC:IQ:RLEN?
//Result: 1564
Manual operation:
See
SYSTem:COMMunicate:RDEVice:OSCilloscope:TCPip
<Address>
Defines the TCPIP address or computer name of the oscilloscope connected to the
R&S
FSW via LAN.
Note
: The IP address is maintained after a PRESET, and is transferred between appli-
cations.
Parameters:
<Address>
computer name or IP address
Example:
SYST:COMM:RDEV:OSC:TCP '192.0.2.0'
Example:
SYST:COMM:RDEV:OSC:TCP 'FSW43-12345'
Manual operation:
See
"TCPIP Address or Computer name"
SYSTem:COMMunicate:RDEVice:OSCilloscope:VDEVice?
Queries whether the connected instrument is supported by the 2
GHz bandwidth
extension option(R&S
FSW-B2000).
For details see the 2
GHz bandwidth extension basics chapter in the R&S
FSW I/Q
Analyzer and I/Q Input User Manual.
Return values:
<State>
ON | 1
Instrument is supported
OFF | 0
Instrument is not supported