Configuration
R&S
®
FSW-K7
96
User Manual 1173.9240.02 ─ 23
"X * Demod
BW"
Sets the step size for the center frequency to a manually defined fac-
tor of the demodulation bandwidth. The "X-Factor" defines the per-
centage of the demodulation bandwidth.
Values between 1 and 100 % in steps of 1 % are allowed. The default
setting is 10 %.
"= Center"
Sets the step size to the value of the center frequency and removes
the coupling of the step size to the demodulation bandwidth. The
used value is indicated in the "Value" field.
"Manual"
Defines a fixed step size for the center frequency. Enter the step size
in the "Value" field.
Remote command:
[SENSe:]FREQuency:CENTer:STEP:LINK
[SENSe:]FREQuency:CENTer:STEP:LINK:FACTor
5.4 Trigger Configuration
Access
: "Overview" > "Trigger"
Triggering means to capture the interesting part of the signal. Choosing the right trigger
type and configuring all trigger settings correctly allows you to detect various incidents
in your demodulated signals.
Optionally, the trigger signal used by the R&S
FSW can be output to a connected
device, and an external trigger signal from a connected device can be used by the
R&S
FSW.
Trigger settings are identical to the base unit, except for the available trigger sources.
Gating is not available for Analog Demodulation measurements.
For background information on trigger settings, trigger output and working with external
triggers, see the R&S
FSW User Manual.
Trigger Configuration