Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
270
User Manual 1173.9240.02 ─ 23
Example:
ADEM:ZOOM:LENG:MODE MAN
Zoom function uses the length defined manually.
Manual operation:
See
[SENSe:]ADEMod<n>:ZOOM:STARt
<Time>
The command selects the start time for the zoomed display of analog-demodulated
measurements in the specified window. The maximum possible value depends on the
measurement time, which is set and can be queried with the
command.
If the zoom function is enabled, the defined number of sweep points are displayed from
the start time specified with this command.
Suffix:
<n>
.
Parameters:
<Time>
Range:
0 s to (measurement time – zoom length)
*RST:
0 s
Example:
ADEM:ZOOM:STAT ON
Switches on the zoom function
ADEM:ZOOM:STAR 500us
Sets the starting point of the display to 500 µs.
Manual operation:
See
[SENSe:]ADEMod<n>:ZOOM[:STATe]
<State>
The command enables or disables the time domain zoom function for the analog-
demodulated measurement data in the specified window.
If the zoom function is enabled, the defined number of sweep points are displayed from
the start time specified with
If the zoom function is disabled, data reduction is used to adapt the measurement
points to the number of points available on the display.
Suffix:
<n>
.
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
ADEM:ZOOM ON
Switches on the zoom function
Manual operation:
See
Configuring the Measurement