Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
187
User Manual 1173.9240.02 ─ 23
Manual operation:
See
11.4 Configuring the Measurement
The following remote commands are required to configure an Analog Demodulation
measurement.
Specific commands:
.................................................................................187
.............................................................................................188
..........................................................................................239
............................................................................................... 242
Configuring the Vertical Axis (Amplitude, Scaling)
................................................243
.................................................................................251
.....................................................................................267
Adjusting Settings Automatically
...........................................................................283
................................................................................ 287
11.4.1 Managing Standard Settings
You can configure the Analog Demodulation application using predefined standard set-
tings. This allows for quick and easy configuration for commonly performed measure-
ments.
For details see
Chapter 5.2, "Configuration According to Digital Standards"
For an overview of predefined standards and settings see
[SENSe:]ADEMod<n>:PRESet[:STANdard]
...................................................................... 187
[SENSe:]ADEMod<n>:PRESet:RESTore
..........................................................................188
[SENSe:]ADEMod<n>:PRESet:STORe
............................................................................ 188
[SENSe:]ADEMod<n>:PRESet[:STANdard]
<Standard>
This command loads a measurement configuration.
Standard definitions are stored in an xml file. The default directory for Analog Demodu-
lation standards is
C:\R_S\INSTR\USER\predefined\AdemodPredefined
.
Suffix:
<n>
.
irrelevant
Parameters:
<Standard>
String containing the file name.
If you have stored the file in a subdirectory of the directory men-
tioned above, you have to include the relative path to the file.
Configuring the Measurement