Configuration
R&S
®
FSW-K7
116
User Manual 1173.9240.02 ─ 23
and f
max
=DBW/2
f
max
and span
min
are specified in the data sheet.
Remote command:
[SENSe:]ADEMod<n>:SPECtrum:SPAN[:MAXimum]
[SENSe:]ADEMod<n>:SPECtrum:SPAN:ZOOM
Demodulation Bandwidth
Defines the demodulation bandwidth of the measurement. The demodulation band-
width determines the sample rate with which the input signal is captured and analyzed.
For recommendations on finding the correct demodulation bandwidth see
For details on the relation between demodulation bandwidth and sample rate refer to
Chapter 4.3, "Sample Rate and Demodulation Bandwidth"
Remote command:
RF Full Span
Sets the span (around the center frequency) of the RF data to be evaluated to the
demodulation bandwidth.
Remote command:
[SENSe:]ADEMod<n>:SPECtrum:SPAN[:MAXimum]
5.7.3 AF Filter
Access
: "Overview" > "Demod Settings" > "AF Filter"
Or
: "Meas Setup" > "Demod" > "AF Filter" tab
The AF filter reduces the evaluated bandwidth of the demodulated signal and can
define a weighting function.
AF filters are only available for AF time domain evaluations.
Demodulation