Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
233
User Manual 1173.9240.02 ─ 23
SYSTem:COMMunicate:RDEVice:GENerator:LINK
........................................................... 234
SYSTem:COMMunicate:RDEVice:GENerator:TYPE
..........................................................234
SYSTem:COMMunicate:TCPip:RDEVice:GENerator:ADDRess
.......................................... 234
SOURce:EXTernal:ROSCillator[:SOURce]
<Source>
This command controls selection of the reference oscillator for the external generator.
If the external reference oscillator is selected, the reference signal must be connected
to the rear panel of the instrument.
Parameters:
<Source>
INTernal
the internal reference is used
EXTernal
the external reference is used; if none is available, an error flag
is displayed in the status bar
*RST:
INT
Example:
SOUR:EXT:ROSC EXT
Switches to external reference oscillator
Manual operation:
See
SYSTem:COMMunicate:GPIB:RDEVice:GENerator:ADDRess
<Number>
Changes the IEC/IEEE-bus address of the external generator.
Parameters:
<Number>
Range:
0 to 30
*RST:
28
Example:
SYST:COMM:GPIB:RDEV:GEN:ADDR 15
Manual operation:
See
"GPIB Address / TCP/IP Address"
SYSTem:COMMunicate:RDEVice:GENerator:INTerface
<Type>
Defines the interface used for the connection to the external generator.
This command is only available if external generator control is active (see
Parameters:
<Type>
GPIB | TCPip
*RST:
GPIB
Example:
SYST:COMM:RDEV:GEN:INT TCP
Manual operation:
See
Configuring the Measurement