![R&S FSW-K7 User Manual Download Page 265](http://html1.mh-extra.com/html/rands/fsw-k7/fsw-k7_user-manual_3082493265.webp)
Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
265
User Manual 1173.9240.02 ─ 23
11.4.7.2
Configuring the Trigger Output
The following commands are required to send the trigger signal to one of the variable
TRIGGER INPUT/OUTPUT connectors on the R&S
FSW.
OUTPut:TRIGger<port>:DIRection
...................................................................................265
........................................................................................ 265
...................................................................................... 266
OUTPut:TRIGger<port>:PULSe:IMMediate
.......................................................................266
OUTPut:TRIGger<port>:PULSe:LENGth
.......................................................................... 266
OUTPut:TRIGger<port>:DIRection
<Direction>
This command selects the trigger direction for trigger ports that serve as an input as
well as an output.
Suffix:
<port>
.
Selects the used trigger port.
2 = trigger port 2 (front panel)
3 = trigger port 3 (rear panel)
Parameters:
<Direction>
INPut
Port works as an input.
OUTPut
Port works as an output.
*RST:
INPut
Manual operation:
See
OUTPut:TRIGger<port>:LEVel
<Level>
This command defines the level of the (TTL compatible) signal generated at the trigger
output.
This command works only if you have selected a user defined output with
.
Suffix:
<port>
.
Selects the trigger port to which the output is sent.
2 = trigger port 2 (front)
3 = trigger port 3 (rear)
Parameters:
<Level>
HIGH
5 V
LOW
0 V
*RST:
LOW
Example:
OUTP:TRIG2:LEV HIGH
Manual operation:
See