Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
231
User Manual 1173.9240.02 ─ 23
Parameters:
<Value>
<numeric value>
*RST:
1
Example:
SOUR:EXT:FREQ:NUM 4
SOUR:EXT:FREQ:DEN 3
Sets a multiplication factor of 4/3, i.e. the transmit frequency of
the generator is 4/3 times the analyzer frequency.
Manual operation:
See
"(Automatic) Source Frequency (Numerator/Denominator/
SOURce:EXTernal:FREQuency[:FACTor]:NUMerator
<Value>
This command defines the numerator of the factor with which the analyzer frequency is
multiplied in order to obtain the transmit frequency of the selected generator.
Select the multiplication factor such that the frequency range of the generator is not
exceeded if the following formula is applied to the start and stop frequency of the ana-
lyzer:
Offset
r
Denominato
Numerator
RF
Freq
Source
Parameters:
<Value>
<numeric value>
*RST:
1
Example:
SOUR:EXT:FREQ:NUM 4"
"SOUR:EXT:FREQ:DEN 3"
Sets a multiplication factor of 4/3, i.e. the transmit frequency of
the generator is 4/3 times the analyzer frequency.
Manual operation:
See
"(Automatic) Source Frequency (Numerator/Denominator/
SOURce:EXTernal:FREQuency:OFFSet
<Offset>
This command defines the frequency offset of the generator with reference to the ana-
lyzer frequency.
Select the offset such that the frequency range of the generator is not exceeded if the
following formula is applied to the start and stop frequency of the analyzer:
Offset
r
Denominato
Numerator
RF
Freq
Source
Parameters:
<Offset>
<numeric value>, specified in Hz, kHz, MHz or GHz, rounded to
the nearest Hz
*RST:
0 Hz
Configuring the Measurement