Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
284
User Manual 1173.9240.02 ─ 23
Manual operation:
See
"Adjusting all Determinable Settings Automatically (Auto
[SENSe:]ADJust:CONFigure:DURation
<Duration>
In order to determine the ideal reference level, the R&S
FSW performs a measurement
on the current input data. This command defines the length of the measurement if
[SENSe:]ADJust:CONFigure:DURation:MODE
is set to
MANual
.
Parameters:
<Duration>
Numeric value in seconds
Range:
0.001 to 16000.0
*RST:
0.001
Default unit: s
Example:
ADJ:CONF:DUR:MODE MAN
Selects manual definition of the measurement length.
ADJ:CONF:LEV:DUR 5ms
Length of the measurement is 5 ms.
Manual operation:
See
"Changing the Automatic Measurement Time (Meastime
[SENSe:]ADJust:CONFigure:DURation:MODE
<Mode>
In order to determine the ideal reference level, the R&S
FSW performs a measurement
on the current input data. This command selects the way the R&S
FSW determines the
length of the measurement .
Parameters:
<Mode>
AUTO
The R&S
FSW determines the measurement length automati-
cally according to the current input data.
MANual
The R&S
FSW uses the measurement length defined by
[SENSe:]ADJust:CONFigure:DURation
*RST:
AUTO
Manual operation:
See
"Resetting the Automatic Measurement Time (Meastime
See
"Changing the Automatic Measurement Time (Meastime
Configuring the Measurement