EM510
TRACe Subsystem
4065.7763.32-01.00
5.122
.
.
:VALue? <trace_name>, <index>, <numeric_value>
Setting an element of a trace.
Note:
Only suppress traces can be set.
Error message:
If the trace name is unknown or not equal to a suppress trace name, an error
-141, "Invalid
character data"
is generated.
Parameters:
<trace_name>
name of the trace to be set as
<Character Data>
SSTART1,SSTOP
<index>
Index of the element within the trace that is to be set. The first
element of a trace has the index 1.
<numeric _value>
frequency value of the element
*RST state:
see
TRACe:DATA
Example:
TRACe:VALue SSTART, 13, 98.550 kHz
.
.
:VALue? <trace_name>, <index>
Query about an element of a trace.
Note:
Only elements of the suppress traces can be queried.
Error message:
If the trace name is unknown or not equal to a suppress trace name, an error
-141 "Invalid
character data "
is generated.
Parameters:
<trace_name>
name of the trace to be read as
<Character Data>
SSTART,SSTOP
<index>
Index of the element within the trace that is to be set. The first
element of a trace has the index 1.
Result:
Frequency value of the element of a trace in Hz
Example:
TRACe:Value? SSTART, 13 -> 98550
Summary of Contents for EM510
Page 4: ......
Page 6: ......
Page 12: ...EM510 Contents 4065 7763 32 01 00 8...
Page 17: ...Characteristics EM510 2 5 4065 7763 32 01 00 Fig 2 1 Front View Fig 2 2 Rear View...
Page 18: ...EM510 Characteristics 4065 7763 32 01 00 2 6 Fig 2 3 Modules...
Page 22: ...EM510 Characteristics 4065 7763 32 01 00 2 10 Fig 2 4 EM510 Detailed Block Diagram...
Page 24: ...EM510 Characteristics 4065 7763 32 01 00 2 12...
Page 30: ...EM510 Installation and Cabling 4065 7763 32 01 00 3 6...
Page 180: ...EM510 Remote Control via LAN 4065 7763 32 01 00 5 144...
Page 183: ...Maintenance and Troubleshooting EM510 6 3 4065 7763 32 01 00...