DIAGnostic Subsystem
EM510
5.27
4065.7763.32-01.00
.
:TEMPerature[<numeric_suffix>]?
Output of the temperature at various measurement points of the device.
Parameters:
The measurement point (1 to 5) is selected through the
<numeric suffix>
.
1 or none
<numeric suffix>
TEMP_LOCAL
temperature on the mainboard
2
TEMP_FPGA
temperature of the FPGA
3
TEMP_PPC
temperature of the Power PC
4
TEMP_ADSECT
temperature of AD converter section
5
TEMP_PRESEL
temperature in the preselection
Result:
Current temperature in °C at the selected test point.
Example:
DIAGnostic:TEMPerature3? -> 37
.
:TPOint[<numeric_suffix>]? <module>
Output of test point voltage of a module. The test point (1 to n
max
) on the particular module is selected
via
<numeric suffix>
:
Parameters:
<module>
information relating to module to be queried:
V1
preselection
n
max
=
3
MB
mainboard
n
max
= 16
Result:
Current test point voltage in mV
Example:
DIAGnostic:TPOint1? V1 -> 1448
Summary of Contents for EM510
Page 4: ......
Page 6: ......
Page 12: ...EM510 Contents 4065 7763 32 01 00 8...
Page 17: ...Characteristics EM510 2 5 4065 7763 32 01 00 Fig 2 1 Front View Fig 2 2 Rear View...
Page 18: ...EM510 Characteristics 4065 7763 32 01 00 2 6 Fig 2 3 Modules...
Page 22: ...EM510 Characteristics 4065 7763 32 01 00 2 10 Fig 2 4 EM510 Detailed Block Diagram...
Page 24: ...EM510 Characteristics 4065 7763 32 01 00 2 12...
Page 30: ...EM510 Installation and Cabling 4065 7763 32 01 00 3 6...
Page 180: ...EM510 Remote Control via LAN 4065 7763 32 01 00 5 144...
Page 183: ...Maintenance and Troubleshooting EM510 6 3 4065 7763 32 01 00...