5.5.3.5 STATus:OPERation:SWEeping Register
This register contains further information about the status of the device. The device is either in normal
reception (fixed frequency) or in one of several scan modes (FSCAN, MSCAN, PSCAN).
The state is determined by the
SENSe:FREQuency:MODE
command with the
CW|FIXed
state being
characterized by deleting bits 3 to 5 in the
STATus:OPERation:SWEeping
register.
Table 5-4: Bit allocation of STATus:OPERation:SWEeping register
Bit No.
Meaning
0
Hold
This bit is set if a FSCAN or MSCAN was interrupted by a fulfilled hold condition.
1
Running up
This bit is set if scanning is to be carried out to increasing frequency values or memory locations.
2
Running down
This bit is set if scanning is to be carried out to decreasing frequency values or memory locations.
3
FSCAN active
This bit is set if FREQ:MODE is set on SWEep.
4
MSCAN active
This bit is set if FREQ:MODE is set on MSCan.
5 - 7
reserved
8
PSCAN active
This bit is set if FREQ:MODE is set on PSCan (software option EM510PS).
9
reserved
Summary of Contents for EM510
Page 4: ......
Page 6: ......
Page 12: ...EM510 Contents 4065 7763 32 01 00 8...
Page 17: ...Characteristics EM510 2 5 4065 7763 32 01 00 Fig 2 1 Front View Fig 2 2 Rear View...
Page 18: ...EM510 Characteristics 4065 7763 32 01 00 2 6 Fig 2 3 Modules...
Page 22: ...EM510 Characteristics 4065 7763 32 01 00 2 10 Fig 2 4 EM510 Detailed Block Diagram...
Page 24: ...EM510 Characteristics 4065 7763 32 01 00 2 12...
Page 30: ...EM510 Installation and Cabling 4065 7763 32 01 00 3 6...
Page 180: ...EM510 Remote Control via LAN 4065 7763 32 01 00 5 144...
Page 183: ...Maintenance and Troubleshooting EM510 6 3 4065 7763 32 01 00...