SENSe Subsystem
:MSCan
EM510
5.89
4065.7763.32-01.00
.
.
:HOLD
.
.
.
:TIME <numeric_value>|MINimum|MAXimum
Setting the hold time during signal-controlled scan continuation. If the signal disappears during the dwell
time, the hold time is started. After completion of the hold time, the scan is continued with the next
frequency even if the dwell time has not yet been completed. If the signal during the hold time exceeds
the squelch threshold, the hold time is reset and the end of the dwell time or the renewed
disappearance of the signal is awaited. The hold time is only important if the control function
"STOP:SIGNal"
(see
SENSe:MSCan:CONTrol
) is switched on.
Parameters:
<numeric_value>
hold time in seconds
MINimum|MAXimum
lowest|highest hold time
*RST state:
0 s
Example:
SWEep:HOLD:TIME 10 ms
.
.
.
:TIME? [MINimum|MAXimum]
Query about the hold time.
Parameters:
none
query about the current hold time
MINimum|MAXimum
query about the lowest|highest hold time
Result:
Hold time in seconds
Example:
MSCan:HOLD:TIME? 0.010
Summary of Contents for EM510
Page 4: ......
Page 6: ......
Page 12: ...EM510 Contents 4065 7763 32 01 00 8...
Page 17: ...Characteristics EM510 2 5 4065 7763 32 01 00 Fig 2 1 Front View Fig 2 2 Rear View...
Page 18: ...EM510 Characteristics 4065 7763 32 01 00 2 6 Fig 2 3 Modules...
Page 22: ...EM510 Characteristics 4065 7763 32 01 00 2 10 Fig 2 4 EM510 Detailed Block Diagram...
Page 24: ...EM510 Characteristics 4065 7763 32 01 00 2 12...
Page 30: ...EM510 Installation and Cabling 4065 7763 32 01 00 3 6...
Page 180: ...EM510 Remote Control via LAN 4065 7763 32 01 00 5 144...
Page 183: ...Maintenance and Troubleshooting EM510 6 3 4065 7763 32 01 00...