*RST
RESET
sets the device to a defined default status. The default setting is indicated in the description
of the commands.
*SRE 0 to 255
SERVICE REQUEST ENABLE
sets the service request enable register to the value indicated. Bit 6
(MSS mask bit) remains 0. This command determines under which conditions a service request is
triggered. Query *SRE? reads the contents of the service request enable register in decimal form.
Bit 6 is always 0.
*STB?
READ STATUS BYTE QUERY
reads out the contents of the status byte in decimal form.
*TRG
TRIGGER
triggers the same actions as the INITiate:CONM[:IMMediate] command.
*TST?
SELFTEST QUERY
triggers the module state test and yields a figure which is to be interpreted as
the bit field:
Result = 0 -> all modules ok
Result
0 -> fault in one or several modules. The information about the possible error can be
queried by means of the
SYSTem:ERRor?
command:
*WAI
WAIT-to-CONTINUE
only permits the servicing of the subsequent commands after all preceding
commands have been executed and all signals have settled (also see "Status Reporting System" on
page 5.128 and "*OPC").
Summary of Contents for EM510
Page 4: ......
Page 6: ......
Page 12: ...EM510 Contents 4065 7763 32 01 00 8...
Page 17: ...Characteristics EM510 2 5 4065 7763 32 01 00 Fig 2 1 Front View Fig 2 2 Rear View...
Page 18: ...EM510 Characteristics 4065 7763 32 01 00 2 6 Fig 2 3 Modules...
Page 22: ...EM510 Characteristics 4065 7763 32 01 00 2 10 Fig 2 4 EM510 Detailed Block Diagram...
Page 24: ...EM510 Characteristics 4065 7763 32 01 00 2 12...
Page 30: ...EM510 Installation and Cabling 4065 7763 32 01 00 3 6...
Page 180: ...EM510 Remote Control via LAN 4065 7763 32 01 00 5 144...
Page 183: ...Maintenance and Troubleshooting EM510 6 3 4065 7763 32 01 00...