![background image](http://html1.mh-extra.com/html/rands/em510/em510_manual_3080638065.webp)
FORMat Subsystem
EM510
5.29
4065.7763.32-01.00
5.3.8 FORMat Subsystem
FORMat
.
:BORDer NORMal|SWAPped
Specifies whether binary data is first to be transferred with low or high byte.
Note:
This command affects only the trace data. For UDP data there is a separate setting option.
Parameters:
NORMal
MSB -> ... -> LSB
SWAPped
LSB ->... -> MSB
*RST state:
NORMal
Example:
FORMat:BORDer SWAPped
.
:BORDer?
Query about the output order for binary data.
Parameters:
none
Result:
NORM, SWAP
Result:
FORMat:BORDer? -> SWAP
.
[:DATA] ASCii|PACKed
Specifies the output format of the following queries:
SENSe:DATA?
TRACe:DATA?
Parameters:
ASCii
output with ASCII format
PACKed
output with internal binary data format
*RST state:
ASCii
Example:
FORMat PACKed