EM510
MEASure Subsystem
4065.7763.32-01.00
5.40
.
:MODE?
Query about the set measuring mode.
Parameters:
none
Result:
CONT
continuous measurement
PER
periodic measurement
Example:
MEASure:MODE? -> PER
.
:TIME <numeric_value>|MINimum|MAXimum|DEFault
Setting the duration for all measurement functions.
Note:
The user is responsible for setting a useful measuring time. Time spans which are too short lead to
faulty measurement results.
Parameters:
<numeric_value>
time span in seconds
MINimum|MAXimum
min/max time span
DEFault
default time span
*RST state:
DEFault
Example:
MEASure:TIME 200 ms
.
:TIME? [MINimum|MAXimum]
Query about the set measuring time.
Parameters:
none
queries the current time span
MINimum|MAXimum
queries the min/max time span
Result:
Time span in seconds; the default time span is indicated by
DEF
Example:
MEASure:TIME? -> 0.2
Summary of Contents for EM510
Page 4: ......
Page 6: ......
Page 12: ...EM510 Contents 4065 7763 32 01 00 8...
Page 17: ...Characteristics EM510 2 5 4065 7763 32 01 00 Fig 2 1 Front View Fig 2 2 Rear View...
Page 18: ...EM510 Characteristics 4065 7763 32 01 00 2 6 Fig 2 3 Modules...
Page 22: ...EM510 Characteristics 4065 7763 32 01 00 2 10 Fig 2 4 EM510 Detailed Block Diagram...
Page 24: ...EM510 Characteristics 4065 7763 32 01 00 2 12...
Page 30: ...EM510 Installation and Cabling 4065 7763 32 01 00 3 6...
Page 180: ...EM510 Remote Control via LAN 4065 7763 32 01 00 5 144...
Page 183: ...Maintenance and Troubleshooting EM510 6 3 4065 7763 32 01 00...