SENSe Subsystem
:FREQuency
EM510
5.73
4065.7763.32-01.00
.
.
.
:SPAN? [MINimum|MAXimum]
Query about the display range of the panorama scan.
Parameters:
none
query about the current display range
MINimum|MAXimum
query about the smallest/widest display range
Result:
Frequency value in Hz
Example:
FREQuency:PSCan:SPAN? -> 2000000
.
.
.
:STARt <numeric_value>|MINimum|MAXimum
Setting of the start frequency of the panorama scan.
Parameters:
<numeric_value>
frequency value
MINimum|MAXimum
setting of the lowest/highest start frequency
*RST state:
1 MHz
Example:
FREQuency:PSCan:STARt 1.5 MHz
.
.
.
:STARt? [MINimum|MAXimum]
Query about the start frequency of the panorama scan.
Parameters:
none
query about the current start frequency
MINimum|MAXimum
query about the lowest/highest start frequency
Result:
Frequency value in Hz
Example:
FREQuency:PSCan:STARt?
1500000