SENSe Subsystem
:SWEep
EM510
5.97
4065.7763.32-01.00
.
.
.
:TIME? [MINimum|MAXimum]
Query about the hold time during signal-controlled scan continuation.
Parameters:
none
query about the current hold time
MINimum|MAXimum
query about the lowest|highest hold time
Result:
Hold time in seconds
Example:
SWEep:HOLD:TIME? 0.010
.
.
:STEP<numeric_value>|MINimum|MAXimum
Setting the frequency stepwidth for the frequency scan.
Parameters:
<numeric_value>
frequency value
MINimum|MAXimum
sets the smallest/biggest frequency stepwidth
*RST state:
1 kHz
Example:
SWEep:STEP 25 kHz
.
.
:STEP? [MINimum|MAXimum]
Query about the frequency stepwidth of a frequency scan
Parameters:
none
query about the current frequency stepwidth
MINimum|MAXimum
query about the smallest|largest frequency stepwidth
Result:
Stepwidth in Hz
Example:
SWEep:STEP? -> 25000