TRACe Subsystem
EM510
5.121
4065.7763.32-01.00
.
:POINts? <trace_name>[,MINimum|MAXimum]
Query about the number of values stored in a trace.
The number of values stored in the suppress traces is always 100. Thus, the MAXimum and MINimum
value is also 100.
The number of IFPAN trace values is always 2047.
Error message:
If the trace name is unknown, an error
-141, "Invalid character data"
will be generated.
Parameters:
<trace_name>
see
TRACe[:DATA]?
no further parameter
query about the current number
MINimum|MAXimum
query about the lowest|highest number
Result:
Number of values
Example:
TRACe:POINts? MTRACE, MAX -> 2048
.
.
:AUTO? <trace_name>
Query about the trace length (auto-adjust).
A 0 (no auto-adjust for trace length) is always output for a suppress trace.
Error message:
If the trace name is unknown, an error
-141, "Invalid character data"
will be generated.
Parameters:
<trace_name>
see
TRACe[:DATA]?
Result:
0
no auto-adjust for trace length
1
auto-adjust for trace length
Example:
TRACe:POINts:AUTO? MTRACE;AUTO? ITRACE -> 1;1
Summary of Contents for EM510
Page 4: ......
Page 6: ......
Page 12: ...EM510 Contents 4065 7763 32 01 00 8...
Page 17: ...Characteristics EM510 2 5 4065 7763 32 01 00 Fig 2 1 Front View Fig 2 2 Rear View...
Page 18: ...EM510 Characteristics 4065 7763 32 01 00 2 6 Fig 2 3 Modules...
Page 22: ...EM510 Characteristics 4065 7763 32 01 00 2 10 Fig 2 4 EM510 Detailed Block Diagram...
Page 24: ...EM510 Characteristics 4065 7763 32 01 00 2 12...
Page 30: ...EM510 Installation and Cabling 4065 7763 32 01 00 3 6...
Page 180: ...EM510 Remote Control via LAN 4065 7763 32 01 00 5 144...
Page 183: ...Maintenance and Troubleshooting EM510 6 3 4065 7763 32 01 00...