PART REPLACEMENT
Replacing the scale card
Cubiscan 325 Service Manual
4 2
removed, carefully lift the boards up and turn them
over. Make sure that the colored load cell
connection wires are not moved or disconnected
during this process.
7. Unscrew the two screws on the bottom of the load
cell summing board. Refer to
8. Remove the scale card and discard it.
9. Locate the replacement scale card and screw it into
place.
10. Screw the load cell summing board in place.
11. Reconnect the load cell summing cable.
12. Replace the load cell summing box lid and screw it in
place.
13. Slide the center cover back into place and tighten
the thumb screws.
14. Power the Cubiscan 325 on.
15. Calibrate the Cubiscan 325 scale. For instructions on
calibrating the scale, see
.
Figure 54
Inside load cell summing box
Figure 55
Side view of load cell summing box
Figure 56
Bottom of load cell summing board
Load cell
summing Cable
Screws
Screws
Scale card
Load cell
summing board
Screws