DIAGNOSTICS
Scale errors
Cubiscan 325 Service Manual
1 4
No zero indicator
Complete the following steps if the zero indicator is not lighting.
1. Make sure that the glass platform is free of all objects.
2. Make sure that there are no objects on or leaning against the glass platform.
3. Make sure the glass platform is securely in place.
4. Verify that no vibration or air movement (fans) is affecting the Cubiscan 325.
5. Check the gap between the scale overload bolts. There should be about a 2mm gap
between the bolts. If needed, adjust the bottom bolt with a wrench.
6. Make sure that all the controller box cables are securely attached.
7. Make sure that all load cell summing box cables are securely attached.
8. Zero the Cubiscan 325 by tapping the
Zero
button that is found on the main screen. When
you zero the Cubiscan 325, make sure the platform is free of all objects.
9. If the zero indicator will still not light, calibrate the scale. For information on calibrating the
scale, see
“Scale Calibration” on page 13
.
10. If, after completing the above steps, the zero indicator is still not lighting, please contact
Cubiscan Technical Assistance at 801.451.7000.