DIAGNOSTICS
Gate movement errors
Cubiscan 325 Service Manual
9
1. When the Cubiscan 325 reboots, make sure that it declares “Cubiscan 325 Ready” after all
the files have been initialized. If it declares “Cubiscan 125 Ready,” the wrong Cubiscan may
be selected in the firmware.
2. If, after completing the above step, the gate cursor is still not functioning, please contact
Cubiscan Technical Assistance at 801.451.7000.
Gate speed warning
The gate speed warning message is displayed if the gate is moved too fast to obtain a
measurement reading. The gate must be moved at a moderate speed to obtain accurate
measurement results.
Irregular gate movement
Complete the following steps if the gate movement is wobbly or uneven.
1. Make sure that the two tracks that control the gate movement are free of debris. Because the
carriers are self-lubricating there should be a lubricant on the tracks. Only the debris should
be removed from the track.
2. Make sure that no cables or objects are impeding the movement of the gate.
3. The carriage assemblies that control gate movement may need to be adjusted or replaced.
For instructions on how to do this, see
“Replacing/Adjusting Carriage Assemblies” on
.
4. If, after completing the above steps, the gate does not move smoothly, please contact
Cubiscan Technical Assistance at 801.451.7000.
Return gate to home error
Complete the following steps if the return gate to home error occurs when the gate is in the
home position (right side of the Cubiscan 325). When the Cubiscan 325 is turned on the gate
must be in the home position.
1. Make sure that the home indicator is lit and the gate is in the home position.
2. Check the proximity sensor light. This is a little red light on top of the sensor. If this light is
out, the cable may be loose or disconnected. Check the cable and make sure it is securely
connected at both ends. If the light still does not light, the proximity sensor may need to be
replaced. If the Cubiscan 325 is off this light will also be off.
3. Check the green encoder light. If this light is out, the cable may be loose or disconnected.
Check the cable and make sure it is securely connected at both ends. If the light still does not
light, the encoder may need to be replaced. If the Cubiscan 325 is off this light will also be
off.