Cubiscan 325 Service Manual
1 5
CHAPTER 2
SCALE CALIBRATION
To calibrate the Cubiscan 325 scale, you will need an official test weight up to 50 pounds (25 kg)
(it is recommended that you calibrate with the maximum weight).
IMPORTANT:
Do not begin scale calibration until you have the test weight. Calibrating without
an accurate known weight (within 0.01 of a lb/kg) can make all future weight readings inaccurate.
When calibrating the scale, the Cubiscan 325 must be stable with no movement of the platform
such as that caused by vibration or air movement.
Scale calibration
Take the following steps to calibrate the Cubiscan 325 scale.
1. At the home screen, tap
MENU
.
2. Tap
CALIBRATE
. Select the
Scale
option if it is not already selected. Make sure the weight
and weight units displayed are correct for the test weight you are using. Tap
Next
to begin
the scale calibration. You can change your units from
lbs
to
kgs
under
MENU > CONFIGURE
> Units
.
Figure 5
Home screen