Maintenance
Removing the Controller Box
CubiScan 110-T
40
3. Slide the controller box forward out of the base. Be careful not to pull on
the attached cables. The cables should be long enough to allow the
controller box to slide out far enough to remove the cover.
4. Disconnect all connectors that are attached to the controller box, as
follows:
•
To remove a sensor connector, press the tab on the connector to
release it, and pull it straight out.
•
To remove a USB connector, pull the connector straight out using
even pressure.
•
To remove the Ethernet cable connector, press the tab on the
connector to release it, and pull it straight out.
•
To remove the load cell connector, turn the screws to loosen the
connector, and pull it straight out.
•
To remove the power connector, take hold of the connector close to
the panel, and pull it straight out using even pressure.
•
To remove a serial cable, loosen the screws (with a screwdriver if
necessary), and pull the cable connector out using even pressure.
5. Verify that all cables have been removed from the controller box, then
pull the box out through the front panel.
Figure 37
Removing the Controller Box