Chapter H - FOOT CONTROL
H-2 Planmeca Compact i
PARTS REPLACEMENT & REPAIR
Technical Manual
d)
Remove the cover assembly by first lifting it upwards at the front and then sliding it backwards
as shown below.
e)
Reassemble in reverse order.
1.2 Replacing chair control springs
a)
Remove the foot control cover as described in section “Removing foot control cover” on page
H-1.
b)
Unscrew the two M4x10 DIN 7991 screws and lift off the chair control assembly.
c)
Unscrew the four M4x8 ULS screws and remove the spring support plate.
d)
Replace the damaged chair control springs with new ones.
NOTE
Always replace both springs.
NOTE
The angle between the spring legs must be closing. If not, bend the spring legs
slightly towards each other.
NOTE
The springs must be positioned in a way, that they can easily overlap each
other (one bends upwards, the other bends downwards).
e)
Reassemble in reverse order.
co
v
e0199.eps
Cover assembly
(1)
(2)
fcntr
lspr
ing0199.eps
M4x10 DIN 7991
M4x8 ULS
Spring support
plate
Chair control
springs
Spring support
Spacer
NOTE THE ORIENTATION
OF THE SPRINGS: THEY MUST
EASILY OVERLAP EACH OTHER
.
Chair control
assembly
Summary of Contents for Compact I
Page 1: ...technical manual 10007462_23 EN...
Page 2: ......
Page 20: ...Chapter A GENERAL TECHNICAL DATA A 12 Planmeca Compact i EMC INFORMATION Technical Manual...
Page 212: ...Chapter H FOOT CONTROL H 18 Planmeca Compact i ADJUSTMENTS Technical Manual...
Page 213: ...Planmeca Compact i I 1 Chapter Technical Manual I DIAGRAMS...
Page 214: ...Chapter I DIAGRAMS I 2 Planmeca Compact i Technical Manual...
Page 216: ......
Page 218: ......
Page 220: ......
Page 222: ......
Page 224: ......
Page 226: ......
Page 228: ......
Page 230: ......
Page 232: ......
Page 234: ......
Page 236: ......
Page 238: ......
Page 240: ......
Page 242: ......
Page 243: ......