Chapter C - SERVICE MODE
Planmeca Compact i C-17
SERVICE MODE DETAILED DESCRIPTION
Technical Manual
2 SERVICE MODE DETAILED DESCRIPTION
The following pages contain detailed information about the service modes, giving useful infor-
mation on how to use and interpret the information available. Reading this chapter gives a
good understanding of the unit’s control system. Used abbreviations are explained in the
beginning of this manual.
Underlined number or values in the heading texts are factory default values.
2.1 MAIN control system related service modes
n.0
Disable programming and limit micromotor speed
(For educational establishment use.) In this mode the programming of the instrument, unit
and/or chair settings can be disabled, and the micromotor power can be limited to 50% of the
maximum speed.
The following settings are possible. The factory default is 0 (zero).
instrument
settings
disabled
unit settings
disabled
chair settings
disabled
micromotor
speed limited
0
1
x
2
x
3
x
x
4
x
5
x
x
6
x
x
7
x
x
x
8
x
9
x
x
10
x
x
11
x
x
x
12
x
x
13
x
x
x
14
x
x
x
15
x
x
x
x
Summary of Contents for Compact I
Page 1: ...technical manual 10007462_23 EN...
Page 2: ......
Page 20: ...Chapter A GENERAL TECHNICAL DATA A 12 Planmeca Compact i EMC INFORMATION Technical Manual...
Page 212: ...Chapter H FOOT CONTROL H 18 Planmeca Compact i ADJUSTMENTS Technical Manual...
Page 213: ...Planmeca Compact i I 1 Chapter Technical Manual I DIAGRAMS...
Page 214: ...Chapter I DIAGRAMS I 2 Planmeca Compact i Technical Manual...
Page 216: ......
Page 218: ......
Page 220: ......
Page 222: ......
Page 224: ......
Page 226: ......
Page 228: ......
Page 230: ......
Page 232: ......
Page 234: ......
Page 236: ......
Page 238: ......
Page 240: ......
Page 242: ......
Page 243: ......