Chapter H - FOOT CONTROL
Planmeca Compact i H-13
ADJUSTMENTS
Technical Manual
f)
Then loosen the upper adjustment screw approx. 1/4 turn so that the foot control pedal can
move freely.
g)
Push the foot control pedal downwards and lock it to this position as follows. Tighten the
angle adjusting screw located on the back of the pedal arm first completely and then loosen
the screw 1/4 turn (see figure on page H-14).
h)
Check that the screws of the sector plate and pedal centering plate are loose.
i)
Place the foot control calibration tool over the foot control so that the attachment screws go
into the Foot control PCB attachment holes. Remove the play of the calibration tool
attachment by pushing the calibration tool towards the pedal end as shown on the figure
below. Make sure that the calibration tool is not tilted.
j)
Tighten the attachment screws gradually as follows. First tighten the screws number one and
two slightly and then screws three and four, and then again screws one and two etc. until the
screws are completely tightened.
k)
Release the pedal from the lower position by loosening the angle adjusting screw.
Fcntr
l5.eps
1/4 round
approx. 1/4 turn
1
3
4
2
Fcntr
l6.eps
Push the pedal downwards and tighten the
angle adjusting screw.
Remove the play.
Summary of Contents for Compact I
Page 1: ...technical manual 10007462_23 EN...
Page 2: ......
Page 20: ...Chapter A GENERAL TECHNICAL DATA A 12 Planmeca Compact i EMC INFORMATION Technical Manual...
Page 212: ...Chapter H FOOT CONTROL H 18 Planmeca Compact i ADJUSTMENTS Technical Manual...
Page 213: ...Planmeca Compact i I 1 Chapter Technical Manual I DIAGRAMS...
Page 214: ...Chapter I DIAGRAMS I 2 Planmeca Compact i Technical Manual...
Page 216: ......
Page 218: ......
Page 220: ......
Page 222: ......
Page 224: ......
Page 226: ......
Page 228: ......
Page 230: ......
Page 232: ......
Page 234: ......
Page 236: ......
Page 238: ......
Page 240: ......
Page 242: ......
Page 243: ......