Chapter C - SERVICE MODE
C-36 Planmeca Compact i
SERVICE MODE DETAILED DESCRIPTION
Technical Manual
n 88
1st vertical segment (counting from right): Scaler electronics activated
2nd: Status LED / Run LED control
3rd: Factory use
4th: Factory use
5th: Not in use
6th: Not in use
7th: Not in use
8th: Not in use
n 89
1st vertical segment (counting from right): Microvac overfill error
2nd: Metasys/Dürr functional
3rd: Metasys/Dürr separator alarm or Dürr VSA float reed relay signal
4th: Factory use
5th: Operation light switch
6th: Cuspidor door switch
7th: Metasys separator beeper
8th: Factory use
n 90
1st vertical segment (counting from right): Scaler electronics selection 1
2nd: Scaler electronics selection 2
3rd: Scaler electronics selection 3
4th: Separator selection 1
5th: Separator selection 2
6th: Separator selection 3
7th: WCS container recognition
8th: Patient glass fill button
n 91
1st vertical segment (counting from right): Not in use
2nd: Not in use
3rd: Backrest motor control
4th: Lift motor control
5th: Chair control
n 92
1st vertical segment (counting from right): Factory use
2nd: Emergency switch
3rd: Factory use
Summary of Contents for Compact I
Page 1: ...technical manual 10007462_23 EN...
Page 2: ......
Page 20: ...Chapter A GENERAL TECHNICAL DATA A 12 Planmeca Compact i EMC INFORMATION Technical Manual...
Page 212: ...Chapter H FOOT CONTROL H 18 Planmeca Compact i ADJUSTMENTS Technical Manual...
Page 213: ...Planmeca Compact i I 1 Chapter Technical Manual I DIAGRAMS...
Page 214: ...Chapter I DIAGRAMS I 2 Planmeca Compact i Technical Manual...
Page 216: ......
Page 218: ......
Page 220: ......
Page 222: ......
Page 224: ......
Page 226: ......
Page 228: ......
Page 230: ......
Page 232: ......
Page 234: ......
Page 236: ......
Page 238: ......
Page 240: ......
Page 242: ......
Page 243: ......