Planmeca Compact i H-7
ADJUSTMENTS
Technical Manual
2 ADJUSTMENTS
2.1 Centering foot control pedal
NOTE
The pedal of the new foot control model is also centered when the sector plate
is calibrated, see section “Calibrating the sector plate and adjusting the angle
of the foot control pedal (new foot control model, from serial number 826740)”
on page H-11.
a)
Remove the foot control cover as described in section “Removing foot control cover” on page
H-1.
b)
Remove the foot control PCB as described in section “Replacing foot control PCB” on page
H-3.
c)
Loosen the two screws which hold the pedal centering plate in position.
d)
By adjusting the position of the pedal centering plate, center the foot control pedal so that it is
positioned exactly in the middle of the foot control casting (A=B).
e)
Tighten the two pedal centering plate screws.
f)
Reassemble the foot control in reverse order.
g)
Calibrate the foot control as described in section “Calibrating foot control” on page H-15.
pedal0199.eps
A
B
A=B
Sector plate
Pedal centering plate
Screws
Axis of the
foot control pedal
Summary of Contents for Compact I
Page 1: ...technical manual 10007462_23 EN...
Page 2: ......
Page 20: ...Chapter A GENERAL TECHNICAL DATA A 12 Planmeca Compact i EMC INFORMATION Technical Manual...
Page 212: ...Chapter H FOOT CONTROL H 18 Planmeca Compact i ADJUSTMENTS Technical Manual...
Page 213: ...Planmeca Compact i I 1 Chapter Technical Manual I DIAGRAMS...
Page 214: ...Chapter I DIAGRAMS I 2 Planmeca Compact i Technical Manual...
Page 216: ......
Page 218: ......
Page 220: ......
Page 222: ......
Page 224: ......
Page 226: ......
Page 228: ......
Page 230: ......
Page 232: ......
Page 234: ......
Page 236: ......
Page 238: ......
Page 240: ......
Page 242: ......
Page 243: ......