Chapter F - ADJUSTMENTS
F-36 Planmeca Compact i
WMS OPERATION
Technical Manual
WCS is started with chip blow key and the bottle has to be filled with cleaning agent. After that
the feeding cycle starts.
Cleaning agent is fed to instruments
City Water
Bottled Water
SW1
off
SW1
on
Feeding cycle:
Instruments
Similar to City water mode
V1
c
V2
c
V3
o
V4
c
V5
c
Pump
not running
Display
countdown timer
Duration
syringe 40 s, rightmost instrument 5
s, following instruments 3 s each
Cleaning agent is fed to cup fill line
City Water
Bottled Water
Feeding cycle:
Cup fill
Similar to City water mode
V1
c
V2
c
V3
o
V4
c
V5
o
Pump
not running
Display
countdown timer 4 s
Duration
4 s