Chapter C - SERVICE MODE
Planmeca Compact i C-43
SERVICE MODE DETAILED DESCRIPTION
Technical Manual
n 126 Show Water management system inputs and outputs
This service mode shows the WMS output and input signals.
1st vertical segment (counting from right): Input, Clean water valve
2nd: Input, Container and pump ON switch
3d: Input, Clean water mode switch
4th: Not in use.
5th: Output, Clean water disable valve drive
6th: Output, Container valve drive
7th: Output, Pump drive
8th: Output, Pump drive enabled
n.127 Water management system water flow adjustment
In this mode the water flow of the Water management system is adjusted.
Check the water flow and adjust the flow by turning the adjustment knob located on the mag-
netic valve/pressure regulator assembly. The water flow can be stopped by pressing the A-
key.
Press the B-key briefly and let the water run for one minute. Pour the water into a glass meas-
ure.
The water flow must be:
Metasys typ1 and Compact, Microvac, no suction: 330 - 400 ml / min.
Dürr and Dürr VSA systems: 400 - 500 ml /min.
n. 128 Fake pressure sensor readings
Dental unit operates totally normally and even tells what pressure levels there is (should be)
on incoming air/water, driver air or spray air/water. For trade shows when using dental unit
without water or air.
0 = normal, 1 = fake pressure readings
n.129 Adjustment of the upper limit of preprogrammed chair
positions
In this service mode it is possible to adjust the upper limit of the preprogrammed chair posi-
tions. Adjustment range is 191…212. The range of the lift motor potentiometer signal is
0…255. In manual operation the maximum value is 240.
The current value from the lift motor potentiometer can be checked from SM 76.
Factory default depends on the chair type, which is shown in the service mode 103, digit 4.
n.130 Disable WCS (0, 1,2)
In this mode the Waterline cleaning system (WCS) operation can be disabled:
0 = WCS in use (factory default)
1 = Starting of WCS disabled, safety features still active
2 = WCS totally disabled, i.e. also safety features are disabled.
Summary of Contents for Compact I
Page 1: ...technical manual 10007462_23 EN...
Page 2: ......
Page 20: ...Chapter A GENERAL TECHNICAL DATA A 12 Planmeca Compact i EMC INFORMATION Technical Manual...
Page 212: ...Chapter H FOOT CONTROL H 18 Planmeca Compact i ADJUSTMENTS Technical Manual...
Page 213: ...Planmeca Compact i I 1 Chapter Technical Manual I DIAGRAMS...
Page 214: ...Chapter I DIAGRAMS I 2 Planmeca Compact i Technical Manual...
Page 216: ......
Page 218: ......
Page 220: ......
Page 222: ......
Page 224: ......
Page 226: ......
Page 228: ......
Page 230: ......
Page 232: ......
Page 234: ......
Page 236: ......
Page 238: ......
Page 240: ......
Page 242: ......
Page 243: ......