Chapter D - HELP & ERROR MESSAGES
Planmeca Compact i D-3
HELP MESSAGES
Technical Manual
HE 16
The operating light is currently dimmed on pur-
pose.
This feature enables easier work with compos-
ite materials, minimizing the risk of operating
light induced curing. This feature is triggered
by picking and returning the curing light to its
holder without starting it. The operating light
returns to normal at the same instance any
control is touched or another instrument
selected. The feature can be disabled, see
page C-32 for details.
HE 17
There is no syringe connected to the instru-
ment console.
The console syringe must be re-connected,
otherwise there is an air-leak from the syringe
QC whenever an instrument or device that
requires air pressure is used.
HE 18
There is no instrument hose connected to this
instrument place (this doriot arm).
The instrument is picked out, but there is no
hose connected at this location.
HE 19
There is no syringe connected to the connec-
tor located on the cuspidor.
The nurse syringe must be re-connected, oth-
erwise there is an air-leak from the nurse
syringe QC whenever an instrument or device
that requires air pressure is used. Install
syringe or disable syringe using service mode
“n.084”.
HE 20
The connector located on the cuspidor allows
only a syringe, other instruments can not be
connected to this connector.
The nurse syringe QC cannot control the cur-
rently connected instrument.
HE 21
The flow of both the spray air and spray water
are programmed to zero (no flow).
Enter the spray programming mode to set a
flow other than zero, see 4.2 “Instrument spray
flow rates” on page B-4 for details.
HE 22
The main air/water valve operation is selected
to be on (n=1) or off (n=0) in the service mode
12 (air)/13 (water). The operation must be
selected to be automatic (n=A).
This is just a reminder that either of the main
valves is set temporarily to be either always on
or off.
HE 23
The automatic/manual chip blow can not be
used when the sterile water function is
selected.
HE 24
The selected setting can not be programmed.
The programming of the selected setting is
disabled. Disable the protection using service
mode “n.000”. See page C-17 for details.
HE 25
Planmeca Compact e help message: The
patient chair cannot be driven into the auto-
matic chair position.
Swivel the patient chair with automatic legrest
before driving the chair to the automatic posi-
tion.
Check that the chair type in n.103 is correct.
(refer to chapter “n 103 Show the dental unit
type (nn)” on page C-39)
HE 26
Compact WE’S help message, should not
appear in Compact i: The chair backrest can-
not be driven down, the seat is too low.
Drive the seat upwards before lowering the
backrest. If this help message is shown while
the patient chair is attached to the unit, it
means that you have forgotten to switch on the
chair’s automatic positions. To clear the mes-
sage, put the operation mode switch on the
cuspidor side of the chair base in the “AUTO”
position (switch up).
Help Code
Meaning
Action or Explanation
Summary of Contents for Compact I
Page 1: ...technical manual 10007462_23 EN...
Page 2: ......
Page 20: ...Chapter A GENERAL TECHNICAL DATA A 12 Planmeca Compact i EMC INFORMATION Technical Manual...
Page 212: ...Chapter H FOOT CONTROL H 18 Planmeca Compact i ADJUSTMENTS Technical Manual...
Page 213: ...Planmeca Compact i I 1 Chapter Technical Manual I DIAGRAMS...
Page 214: ...Chapter I DIAGRAMS I 2 Planmeca Compact i Technical Manual...
Page 216: ......
Page 218: ......
Page 220: ......
Page 222: ......
Page 224: ......
Page 226: ......
Page 228: ......
Page 230: ......
Page 232: ......
Page 234: ......
Page 236: ......
Page 238: ......
Page 240: ......
Page 242: ......
Page 243: ......